Home Precision Environment Aerosol Conditioning DIL 541/C Controlled Dilution System
DIL 541/C Aerosol Dilution System product image thumbnail
Topas GmbH Aerosol Dilution Systems

DIL 541/C Controlled Dilution System

Self-adjusting 1:10 aerosol dilution system for 28.3 L/min particle-counter flow, with automatic ratio control and digital status communication.

Dilution factor
1:10
Aerosol flow
28.3 L/min
Control
Automatic, controlled
Counterpressure
±20 kPa
1:10 28.3 L/min Automatic control Digital status
VDI 3491-6 ISO 14644-3

Model Overview

Model Configuration Primary use Selection note
DIL 541/C 1:10; 28.3 L/min Self-adjusting 1:10 aerosol dilution system for 28 Select DIL 541/C for automatically controlled 1:10 dilution with a 28.3 L/min cleanroom particle counter.

Key Features

Automatic dilution-ratio control

The dilution ratio is automatically adjusted and monitored to maintain defined aerosol conditioning during changing operating conditions.

Digital status communication

Supports status communication for integration with particle counters, test benches and supervised measurement workflows.

Closed aerosol path

Directs the complete intake flow of the connected particle counter from the sampling point through the dilution system.

Isokinetic capillary sampling

The fluidic design minimises changes to the aerosol particle-size distribution during splitting and recombination of the flows.

Positive and negative pressure operation

Designed for operation under overpressure or underpressure within the published counterpressure limits.

Cascadable configuration

Can be connected in series when a larger overall dilution factor is required for the measurement task.

Technical Specifications

Parameter Unit Value / Description
CONDITIONING
Purpose Dilution of aerosols
Sample condition, inlet Non-diluted aerosol
Sample condition, outlet Diluted aerosol
Dilution factor 10
Aerosol volumetric flow rate L/min 28.3
Setting method Automatic, controlled
CONNECTIONS & OPERATING CONDITIONS
Power supply 12 V DC, 0.1 A; optional battery
Hose connector mm Ø 8, optional Ø 10
Pressure drop kPa 0.8–1.3
Environmental temperature °C 10–50
Maximum counterpressure kPa ±20
PHYSICAL
Dimensions (W × H × D) mm 110 × 160 × 300
Weight kg 1.5

Specifications follow the current Topas GmbH product information reviewed for this page. Confirm the final configuration, accessories and calibration documentation at quotation stage.

Applications

Verification of cleanroom test aerosols

Reduces high upstream challenge-aerosol concentrations to a range suitable for optical particle counters during cleanroom and clean-air-device testing.

Laminar-flow and safety-cabinet acceptance

Supports acceptance measurements on laminar-flow boxes, work benches, safety cabinets and smaller cleanroom installations.

Filter separation-efficiency measurement

Extends the usable concentration range of particle counters in upstream and downstream filter-efficiency measurements.

Highly concentrated aerosol analysis

Conditions concentrated aerosol samples before particle counting to reduce coincidence-related measurement errors.

Aerosol research

Provides a defined dilution stage for experimental aerosol generation, transport and particle-size-distribution studies.

Measurement-range extension

Allows particle-measurement instruments to analyse aerosol concentrations that would otherwise exceed their operating range.

Similar Products